For full functionality of this site it is necessary to enable JavaScript.

GW instek GLC-9000 Leakage Current Tester

 

General

Display

320 x 240 Color TFT LCD

 

Operation

Touch panel

 

Measurement Device

9 kinds

DC

Ranges

50uA/500uA/5mA/25mA

 

Resolution

0.01uA/0.1uA/1uA/10uA

 

Accuracy

±1.0% fs/±(0.2% rdg+3dgt)

AC/AC+DC

Ranges

50uA/500uA/5mA/25mA

 

Resolution

0.01uA/0.1uA/1uA/10uA

 

Accuracy

±2.0% fs/±(2.0% rdg+6dgt) -- 10Hz<f100kHz

 

 

±2.0% fs/±(2.0% rdg+10dgt) -- 100kHz<f1MHz

AC peak

Ranges

500uA/1mA/10mA/75mA

 

Resolution

0.1uA/1uA/10uA/100uA

 

Accuracy

±4.0%fs/±2.5%fs/±(2.0% rdg+6dgt)
-- 20Hz<f
1kHz

 

 

±5.0%fs/±5.0%fs/±(5.0% rdg+10dgt)
-- 1kHz<f
10kHz

Meter Mode

AC/DC/AC+DC

10.0~300.0V

 

AC peak

15.0~430.0V

 

SELV function

Available

EUT

Voltage range

85V ~ 300V

 

Current range

0.5A ~ 10A

Memory

Empty sets

30 sets

 

Preset sets

50 sets (conform to IEC 60990)

Interfaces

Front panel

USB host

 

Rear panel

RS-232C/GPIB/EXT I/O/USB device

Power Source

For GLC-9000

AC 100V/120V/220V/230V±10%50/60Hz

 

For EUT

AC 85V250V50/60Hz

Dimensions

 

330 (W) x 150 (H) x 350 (D) mm

Weight

 

Approx. 5kg

 

Details
  • Quality Engagement
  • Easy change and return
  • Delivery Avaliable
  • Favorable payment

Apply your mail to get promotion information