For full functionality of this site it is necessary to enable JavaScript.

HIOKI CT6844 AC / DC 電流プローブ

Rated current: 500 A AC/DC
Max. allowable input: 720 A DC continuous (requires derating at frequency or temperature)
Frequency characteristics: DC to 200 kHz
Phase: DC to 200 kHz
Amplitude and phase accuracy: 
- DC ±0.3 % rdg. ±0.02 % f.s. (Phase: Not defined)
- DC < f ≤ 100 Hz ±0.3 % rdg. ±0.01 % f.s. (Phase: ±0.1 deg)
- Defined to 200 kHz
Power consumption : 7 VA max. (at 1000 A/55 Hz, ±12 V power requirement)
Output voltage: 4 mV/A (voltage output with the Sensor Unit 9555-10, use with a device having a 1 MΩ input resistance or higher)
Core diameter: φ 20 mm (1.79 in)
Power supply: DC ±11 V to ±15 V (Power suppled via the 9555-10, which supports 100 to 240 V AC)
Dimensions and mass: 153 mm (6.02 in)W × 67 mm (2.64 in)H × 25 mm (0.98 in)D, cord length: 3 m (9.84 ft), 380 g (13.4 oz)
Accessories: Instruction manual ×1, Mark bands ×6, Carrying Case

Details
Rated current 500 A AC/DC
Max. allowable input 720 A DC continuous (requires derating at frequency or temperature)
Frequency characteristics DC to 200 kHz
Phase: DC to 200 kHz
Amplitude and phase accuracy DC ±0.3 % rdg. ±0.02 % f.s. (Phase: Not defined)
DC < f ≤ 100 Hz ±0.3 % rdg. ±0.01 % f.s. (Phase: ±0.1 deg)
Defined to 200 kHz
Power consumption 7 VA max. (at 1000 A/55 Hz, ±12 V power requirement)
Output voltage 4 mV/A (voltage output with the Sensor Unit 9555-10, use with a device having a 1 MΩ input resistance or higher)
Core diameter φ 20 mm (1.79 in)
Operating temperature, humidity -40 °C to +85 °C (-40 °F to 185 °F), 80% rh or less (with no condensation)
Power supply DC ±11 V to ±15 V (Power suppled via the 9555-10, which supports 100 to 240 V AC)
Dimensions and mass 153 mm (6.02 in)W × 67 mm (2.64 in)H × 25 mm (0.98 in)D, cord length: 3 m (9.84 ft), 380 g (13.4 oz)
Accessories Instruction manual ×1, Mark bands ×6, Carrying Case

 